The combined effect of the power and the carrier gas flow rate on excitation temperature, electron number density, ionic-to-atomic line intensity ratios and departure from local thermal equilibrium has been studied in an axially-viewed plasma with introduction of different amounts of wet aerosols, partially desolvated aerosols or dry vapours, according to empirical modelling and experimental design methods.