锤子
Analysis Example by Vacuum Model FT/IR (1) Measurement of Silicon Wafer (1/1) No. 220TR0164-E Due to the high refractive index of silicon wafer samples, there is a correspondingly high reflectance from the sample. If the sample is placed directly in the sample compartment, vertical to the incident source beam, the reflection of the source returns to the interferometer, is reflected by the beamsplitter and then illuminates the sample again. As a result, the sample pathlength of the sample measurement is different from the background measurement and excessive noise is observed due to the atmospheric water vapor, even with a purged instrument. While it is possible to adjust the angle of the sample with respect to the incident source illumination, using an evacuated FT-IR instrument allows measurement of the sample vertical to the incident illumination without
打开失败或需在电脑查看,请在电脑上的资料中心栏目,点击"我的下载"。建议使用手机自带浏览器。