A methodology using capillary electrophoresis with tandem mass spectrometry detection (CE-MS/MS) was developed, and applied to determine the cocaine content of seized drugs. The cocaine separation among other concomitants by capillary electrophoresis was achieved in just 4.0 minutes using 30 mM ammonium formate, pH 6.5, as a background electrolyte (BGE). The correlation coefficient of the calibration curve in the range of 0.02 to 1 μg/mL was upper 0.999, and the limits of detection (LOD) and limits of quantification (LOQ) were 0.003 μg/mL and 0.01 μg/mL, respectively. The relative standard deviation repeatability (intraday and interday) was less than 6.5% (peak area). The analysis of seized suspected illegal drugs revealed the presence of cocaine in 93% of samples (n = 11), in amounts ranging from 0.04 to 429 mg/g, showing a high variability of concentrations of cocaine. The user can change the provider and, due to the large variability of concentration, be lead to an accidental overdose

方案下载
配置单
方案详情

A methodology using capillary electrophoresis with tandem mass spectrometry detection (CE-MS/MS) was developed, and applied to determine the cocaine content of seized drugs. The cocaine separation among other concomitants by capillary electrophoresis was achieved in just 4.0 minutes using 30 mM ammonium formate, pH 6.5, as a background electrolyte (BGE). The correlation coefficient of the calibration curve in the range of 0.02 to 1 μg/mL was upper 0.999, and the limits of detection (LOD) and limits of quantification (LOQ) were 0.003 μg/mL and 0.01 μg/mL, respectively. The relative standard deviation repeatability (intraday and interday) was less than 6.5% (peak area). The analysis of seized suspected illegal drugs revealed the presence of cocaine in 93% of samples (n = 11), in amounts ranging from 0.04 to 429 mg/g, showing a high variability of concentrations of cocaine. The user can change the provider and, due to the large variability of concentration, be lead to an accidental overdose

上一篇 使用 Agilent 5800 ICP-OES 测定固态 电解质锂镧锆钽氧 (LLZTO) 中的 主量元素
下一篇 使用 Agilent Cary 630 ATR-FTIR 光谱仪测定婴儿米粉中的蔗糖浓度

文献贡献者

相关仪器 更多
相关方案
更多

相关产品

当前位置: 安捷伦 方案 采用 CE-MS/MS 检测查获毒品中的可卡因

关注

拨打电话

留言咨询