Models | ANIR 0.9-1.7 | ANIR 0.9-2.6 | AMIR (Link) | HT-NIR (link) |
Spectral Range [nm] | 900-1700 | 900-2600 | 2000-6000 | 900-1700 |
Signal-to-Noise (single measurement full detector dynamic used, Ratio between baseline and white light source measurement) | > 1000:1 | > 1000:1 | > 1000:1 |
> 300:1 Ideal for diffuse samples |
Detector Type | InGaAs | InGaAs | Cooled (Hg,Te, Cd)Zn from Vigo systems |
InGaAs 512 (pixel detector array) |
Features | ARCspectro ANIR |
Spectral resolution in wavenumber: | 8cm-1 |
Spectral resolution in wavelength: at 1000nm: at 1700nm: at 2600nm: |
1.5 nm 4 nm 9 nm |
Wavelength Stability | < 0.1nm |
Absolute Wavelength accuracy | < 0.5nm |
Long term Intensity stability | < 2% (1100-2500nm) over 10 hours |
Minimum scan time | 1 s |
Effective Measurement time | 3 s |
Optical Fiber Connector | SMA 905 |
Optical entrance (optical fiber core) | 0.6 mm |
Sampling Rate | 125 kHz |
A/D Converter | 24 bit |
Operating Voltage | 5V (USB powered). Except for MIR version (TE cooling need external powering). |
Operating temperature | 10°C-30°C |
Entrance aperture | 1 mm (or fiber diameter) |
Communication Interface | USB 2.0 |
Software Interface | Windows XP |
Product Dimensions (including electronics) | 10cm x 15cm x7cm |
Product Weight | 850 g |