1. Product 
YL Methylparaben/ Phenoxyethanol Analyzer

2. Purpose
YL Methylparaben Analyzer efficiently analyzes Methylparaben and Phenoxyethanol which are very commonly
used as preservatives for cosmetics, drugs and other skincare products but they may cause toxicity at
higher concentration to human body. 
For they have an absorbance at UV/Vis wavelength range, these are usually analyzed by HPLC after
extracted by organic solvents such as Methanol or DMF.
Our dedicated analyzer for Methylparaben/Phenoxyethanol supplies a photo diode array detector to 
provide accurate results separating all other impurities in the sample.  

3. Analysis of Methylparaben/ Phenoxyethanol 
1) Principle of analysis
• Both Methylparaben and Phenoxyethanol have an absorbance at UV/Vis wavelength range which means
 these can be analyzed at the maximum absorption.

2) Caution
• The maximum absorptio(λmax) differs depending on compounds. For the maximum absorption of Methylparaben is 257 nm 
and Phenoxyethanols is 271 nm, if analyzing at a single wavelength by UV/Vis detector,one of them loses sensitivity which means the analysis can be proceeded efficiently.

• This analysis has to be done in a gradient mode to have good resolution of both Methylparaben and
 Phenoxyethanol. It is very important to set the gradient to mode for retention time (RT) of these 2 
compounds not to get overlapped and remove the other compounds in the sample that will be also eluted. 
• Even though the analysis condition is well set to get a different RT for each compound after making
 standard samples, there could be some other compounds in the sample that have same RT as standard 
samples’.

 Because UV/Vis detector recognizes the wrong compounds as standard samples if the RT is same, PDA 
detector is recommended for an accurate qualitative analysis to check the exact spectrum. 

If RT is same but spectrum is not, it means that this could be a different compound. Specially,
 Parabens are usually analyzed at the low concentration and it appears occasionally even though it’s not added. These can be residuals from some kind of extractions, the spectrum that can see whether 
there Parabens are or not is essentially needed.

3) Sample & standard preparation 


• Skins & Lotions
Sample 2 ~ 3 g + MeOH 10 mL --à Hand-shaking for dispersion
                           Ultrasonication (20 min)
Mixed Solvent / 50 mL MeOH Mass UP

 Filtering the supernatant of extracted solution to 0.45 um filter

Analysis by HPLC/PDA 

• Creams & Sun Creams
Sample 2 ~ 3 g + DMF 10 mL --à Hand-shaking for dispersion
                           Ultrasonication (10 min)
Mixed Solvent / 50 mL MeOH Mass UP
                          Ultrasonication (20 min)
Filtering the supernatant of extracted solution to 0.45 um filter

Analysis by HPLC/PDA 

4) Results

[Chromatogram]



[Comparison of PDA spectrum depending on compounds]


It is confirmed that Peak 1 and Phenoxy ethanol are same compounds for their spectrum patterns are
 matched but spectrum patterns of Peak 2 and Methyl paraben are not same which means these are not 
the same compounds but have same RT.



4. Features
1) YL Methylparaben Analyzer provides all required equipments and accessories for your convenience. 
2) This system is built to provide an exceptional cost/performance ratio.
3) This system delivers exact application supports according to corroborative analysis methods.

5. Application
1) Row material standards and methods set-up of cosmetics


 