PM1401MB | PM1401GNB | |||
探测器 | ||||
γ | CsI(Tl) | CsI(Tl) | ||
中子 | He-3 | |||
灵敏度 | ||||
相对于137Сs |
100 (s-1)/(μSv/h) (1.0 (s-1)/(μR/h)) |
100 (s-1)/(μSv/h) (1.0 (s-1)/(μR/h)) |
||
相对于241Am |
100 (s-1)/(μSv/h) (1.0 (s-1)/(μR/h)) |
200 (s-1)/(μSv/h) (2.0 (s-1)/(μR/h)) |
||
能量范围 | ||||
γ | 0.033 – 3.0 MeV | 0.033 – 3.0 MeV | ||
中子 | 从热能到 14.0 MeV | |||
剂量率 | ||||
γ |
0.01 – 99.99 μSv/h (1 – 9999 μR/h) |
0.01 – 99.99 μSv/h (1 – 9999 μR/h) |
||
中子 | 1 - 999 s-1 | |||
剂量 | ||||
准确率 |
±30% (in range 0.1 – 70 μSv/h (10 – 7000 μR/h) (at 137Cs) ) |
±30% (in range 0.1 – 70 μSv/h (10 – 7000 μR/h) (at 137Cs)) |
||
响应时间 | 0.25 s | 0.25 s | ||
放射性核素鉴别 | ||||
特种核材料 (SNM) | 233U, 235U, 237Np, Pu | 233U, 235U, 237Np, Pu | ||
医疗放射性核素 |
18F, 67Ga, 51Cr, 75Se, 89Sr, 99Mo, 99mTc,103Pd, 111In, 123I, 131I,153Sm, 201Ti, 133Xe |
18F, 67Ga, 51Cr, 75Se, 89Sr, 99Mo, 99mTc,103Pd, 111In, 123I, 131I,153Sm, 201Ti, 133Xe |
||
自然产生的放射性核素 (NORM) |
40K, 226Ra, 232Th and daughters, 238U and daughters |
40K, 226Ra, 232Th and daughters, 238U and daughters |
||
工业放射性核素 |
57Co, 60Co, 133Ba, 137Cs, 192Ir, 226Ra, 241Am |
57Co, 60Co, 133Ba, 137Cs, 192Ir, 226Ra, 241Am |
||
达标合格 |
ITRAP/IAEA 需求, ANSI N42.32, ANSI N42.33 (1), ANSI N42.34, IEC 62401 |
ITRAP/IAEA r需求, ANSI N42.32, ANSI N42.33 (1), ANSI N42.34, IEC 62401 |
||
报警方式 | 视觉,声音,振动 | 视觉,声音,振动 | ||
数据记录 | 1000 | 1000 | ||
环境等级防护 | IP65 | IP65 | ||
跌落地板测试 | 1.5 m (4.9 ft) | 1.5 m (4.9 ft) | ||
电池供应 | 一款 AA电池 | 一款 AA电池 | ||
电池寿命 | 1000 小时 | 1000 小时 | ||
操作温度 |
-30°C — 50°C (-22°F — 122°F) |
-30°C— 50°C (-22°F — 122°F) |
||
尺寸 (无皮套) |
103 x 57 x 32 mm (4'' x 2¼'' x 1¼'') |
194 x 57 x 32 mm (7 5/8'' x 2¼'' x 1¼'') |
||
重量 | ||||
(无皮套) | 270 g ( 9.52 oz) | 430 g (15.2 oz) | ||
低电量警报 | LCD | LCD | ||
超载报警值 | ||||
γ | OL | OL | ||
中子 | 999 | |||
通讯方式 | 红外数据通讯,蓝牙 | 红外数据通讯,蓝牙 |