★ 技术参数
☆ 温度范围:
A型:-20℃~+100℃(+150℃)
B型:-40℃~+100℃(+150℃)
C型:-60℃~+100℃(+150℃)
D型:-70℃~+100℃(+150℃)
☆ 温度波动度:±0.5℃
☆ 温度均匀度:±2.0℃
☆ 湿度范围:30%~98%R.H
☆ 湿度误差:+2/-3%R.H
★ 型号及工作室尺寸(D×W×H)(mm)
GDJ-100 400×450×550
GDJ-150 500×500×600
GDJ-225 500×600×750
GDJ-400 600×800×850
GDJ-800 800×1000×1000
GDJ-1000 1000×1000×1000
GDJ-1500 1500×1000×1000
BR-PV-THD Thermal Cycling, Humidity-Freeze, Damp-Heat Tester
Thermal Cycling Test
Implementation of standards: IEC 61215:2005 & Ed.3, IEC 61646:2008 “TC200 thermal cycling test, TC50 thermal cycling test”, and UL 1703-2008 “temperature cycling test”.
1. IEC 61215 Ed.3, IEC 61646:2008 test condition:
The temperature of modules between -40°C±2°C and +85°C±2°C cycling, the rate of change of temperature between the low and high extremes shall not exceed 100°C/h and the module temperature shall remain stable at each extreme for a period of at least 10 min. The cycle time shall not exceed 6h.
2. UL 1703-2008 test condition:
The temperature of modules between -40°C±2°C and +90°C±2°C cycling, the instantaneous rate of temperature change of the test chamber with respect to time is not to be greater than 120°C/h.
Humidity-Freeze Test
Implementation of standards:
IEC 61215 Ed.3, IEC 61646:2008 “humidity-freeze”, UL 1703-2008 “humidity test”
The temperature of modules between -40°C±2°C and +85°C±2°C cycling.
Relative humidity 85%RH±5% at 85°C in IEC 61215 and 61646, but UL 1703 Control in 85%RH±2.5% at 85°C, allowable deviation range different.
Damp-Heat Test
Implementation of standards: IEC 61215:2005 & Ed.3, IEC 61646:2008 “damp-heat test”
Test temperature 85°C ± 2°C, Relative humidity 85% ± 5%, Test duration 1000h
Electrical continuity, Insulation continuity, Grounding continuity test system:
Implementation of standards: IEC 61215:2005 & Ed.3, IEC 61646:2008 “electrical continuity for TC200 thermal cycling test and humidity-freeze test” condition, and UL 1703-2008 “Insulation continuity and grounding continuity” condition.
DC power current / voltage output: ≥10A / 50V
Module type | Imp | TC 200 Thermal Cycling | Humidity-Freeze |
1.58×0.808m monocrystalline | 5.20A | Provide Im over 25°C Provide 0.1~0.5A below 25°C | Provide current during the whole test |
2×1m polycrystalline | 7.66A | Provide Im over 25°C Provide 0.1~0.5A below 25°C | Provide current during the whole test |
1.4×1.1m Thin-film | 1.66A | Provide current during the whole test | Provide current during the whole test |
2.6×2.2m Thin-film | 2.66A | Provide current during the whole test | Provide current during the whole test |
用户单位 | 采购时间 |
---|---|
Zhejiang Sunflower Light Energy Science & Technology LLC | 2010-09-01 |
Hengdian Group DMEGC magnetic Co.,Ltd. - Photovoltaic Laboratory | 2010-06-01 |
Chengdu Guibao Science & Technology Co.,Ltd. | 2010-06-01 |
ShanXi LuAn Photovoltaic Technology Co.,Ltd. | 2010-06-01 |
Chinese Academy of Sciences Key Laboratory - New Thin-film Solar Cell | 2008-06-01 |
Worldwide Energy and Manufacturing USA, Inc. ( Chinese factory ) | 2007-06-01 |
相关产品