A 20-30 times more sensitive method than traditional GC- ECD has been developed and validated for the determination of PCBs in soil and water. Methods of analysis were evaluated through determination of detection limits (down to sub ppb level), quanti?cation limits (down to few ppb), linearity and repeatability. Validation was completed by the determination of PCBs extraction recoveries obtained using Liquid/Liquid and ASE extraction. LV-SL has proven to be a reliable tool for trace analysis in environmental applications, being robust against “very dirty” samples like soil extracts. LV-SL is theeasiest large volume injection GC technique, since it doesnot require any special hardware or tedious tuning of operating parameters. Full automation is provided through the dedicated LV-SL Assistant software.
Environmental legislations approved by federal agencies
like US EPA, or by European Commission Directives [1,2],
aim at the most severe reduction of the PCB levels in
water and soils (< 1 ng/kg in the original sample). This
implies, for a PCB analytical method to be validated, the
achievement of very low detection limits (< 0.5 ng/kg in
the extract).
The identification and quantitation of PCBs is most
often accomplished by gas chromatographic techniques
coupled with Electron Capture Detector (ECD), which has
selective sensitivity to chlorinated compounds. Method
sensitivity can be further amplified using large sample
volume injection into the GC [3]. Among the different
techniques used for large volume injection, the recently
introduced LV-SL [4] has the following advantages: (1) It
is simpler because it allows injections, up to 50 µL, in a
conventional split/splitless injector without any special
tuning of operating parameters, and (2) it is robust when
analyzing sample by-products or contaminants, such as
those found in food matrices or soil extracts. It is also
suitable for the analysis of labile pesticides, and it is
compatible with any GC detector [5].
In this method, the application of Large Volume
Splitless (LV-SL) to GC-ECD is validated for the
determination of polychlorinated biphenyls (PCBs) in soil
and water samples. GC-ECD analyses of 20-30 µL of (1)
standard Aroclor mixtures and (2) Liquid/Liquid or
Accelerated Solvent Extraction (ASE) extracts of spiked
water/soil matrices are reported. Injections were
performed with the TriPlus Autosampler; GC oven, and
conventional SSL injector conditions were set-up with
dedicated software; the analytical method was validated
through determination of detection limits, linearity,
extraction recoveries, and repeatability.
A 20-30 times more sensitive method than traditional GC-
ECD has been developed and validated for the
determination of PCBs in soil and water. Methods of
analysis were evaluated through determination of
detection limits (down to sub ppb level), quantification
limits (down to few ppb), linearity and repeatability.
Validation was completed by the determination of PCBs
extraction recoveries obtained using Liquid/Liquid and
ASE extraction.
LV-SL has proven to be a reliable tool for trace
analysis in environmental applications, being robust
against “very dirty” samples like soil extracts. LV-SL is the
easiest large volume injection GC technique, since it does
not require any special hardware or tedious tuning of
operating parameters. Full automation is provided
through the dedicated LV-SL Assistant software.
高压离子色谱Inuvion快速测定自来水中的常见阴离子
高压离子色谱Inuvion快速测定生活饮用水中的亚氯酸 盐、溴酸盐、氯酸盐、二氯乙酸和三氯乙酸
Thermo Fisher Scientific iCAP TQe ICP-MS/MS测定土壤和沉积物多种元素
相关产品
赛默飞Inuvion离子色谱系统
TSQ 9610 三重四极杆 GC-MS/MS
赛默飞气相色谱仪Trace 1600
ISQ 7610™单四极杆 GC-MS
Vanquish Neo UHPLC 系统
赛默飞TSQ Quantis Plus 三重四极杆质谱仪
赛默飞TSQ Altis Plus 三重四极杆质谱仪
赛默飞TSQ Fortis Plus三重四极杆质谱仪
Orbitrap Exploris MX 质量检测器
Orbitrap IQ-X Tribrid 质谱仪
赛默飞三重四极杆 LC-MS质谱系统TSQ Altis(LC-QQQ)
Orbitrap Exploris GC 气质联用仪
Orbitrap Exploris GC 240
新一代Orbitrap Exploris 240质谱仪
Thermo Scientific™ Orbitrap Exploris 120
关注
拨打电话
留言咨询