瑞士万通 6.2325.000 PH 玻璃电极说明书

2017-10-30 17:11  下载量:10

资料摘要

资料下载

6.2325.000 pHit kit Care kit for pH glass electrodes with liquid electrolyte The following steps should be carried out at regular intervals: 1. Close the filler opening. 2. In the case of electrodes with flexible ground-joint diaphragm: Loosen the ground-joint diaphragm. 3. Immerse the electrode in cleaning solution deep enough to cover the diaphragm. 4. Allow the electrode to react in cleaning solution at room temperature for approx. 4 h. 5. Empty the electrolyte chamber. 6. Rinse the electrolyte chamber with electrolyte several times. 7. Immerse the electrode in storage solution for conditioning for at least 2 h.