大家的实验室里有没有定时器?都是什么样子的?
实验室用的小定时器去哪买?见下面的图片。http://ng1.17img.cn/bbsfiles/images/2014/02/201402061516_489559_2826867_3.jpg
车间化验室的一只机械定时器出现不闹铃故障。这个定时器全身没有一颗螺丝,想办法拆开看看,是振铃块脱落,修复后恢复正常工作。下面是拆修过程。外观,鸡蛋型,不锈钢外壳,比较漂亮。定时范围60分钟:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719382825_5776_1807987_3.jpg!w690x517.jpg[/img]全身没有一颗螺丝,不知道从何处下手:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719386706_1103_1807987_3.jpg!w690x517.jpg[/img]试着握紧上、下部分,拉开:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719390706_5584_1807987_3.jpg!w690x517.jpg[/img]分离开后,内部掉出一个小铁块:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719390785_8300_1807987_3.jpg!w690x517.jpg[/img]仔细端详,是从箭头处脱落的,这个地方是振铃臂杆,位置狭小,不容易装回去:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719394226_4955_1807987_3.jpg!w690x517.jpg[/img]机芯上的三颗固定螺丝不能松开!松开后机芯会散架,再装还原就比较困难了。箭头所指四个孔内,也没有固定螺丝,机芯取不出来:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719401477_3483_1807987_3.jpg!w690x517.jpg[/img]准备撬出机芯,先在缺口处做一个记号:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719402407_2039_1807987_3.jpg!w690x517.jpg[/img]用小刀在不锈钢外壳与塑料内衬之间划划,底部粘得太牢,取不出机芯。放弃这个方法:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719406438_1004_1807987_3.jpg!w690x517.jpg[/img]使用平口改刀沿塑料内衬圆周,用力向外撬机芯铁壳:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719414799_8452_1807987_3.jpg!w690x517.jpg[/img]“咔”的一声,分离开机芯,原来塑料内衬中心有一颗固定螺丝,被暴力分开(看来,确实应该先取出塑料内衬,卸下中心固定螺丝,就可取下机芯。无奈塑料内衬与不锈钢外壳粘死,坑人的工艺!看来是一次性用品):[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011719375892_8561_1807987_3.jpg!w690x517.jpg[/img]在机芯上作好位置画痕记号:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721127382_2698_1807987_3.jpg!w690x517.jpg[/img]卸下机芯保护底壳上的三颗固定螺丝:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721132583_1508_1807987_3.jpg!w690x517.jpg[/img]取下机芯:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721139651_6528_1807987_3.jpg!w690x517.jpg[/img]这一面是走时发条和闹铃发条:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721139915_5731_1807987_3.jpg!w690x517.jpg[/img]去掉保护壳后,振铃臂杆处位置宽敞,容易将振铃铁块安装回去:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721143836_7496_1807987_3.jpg!w690x517.jpg[/img]滴一滴502,将振铃铁块牢牢地粘在振铃臂杆上:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721150542_8806_1807987_3.jpg!w690x517.jpg[/img][img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721150207_8681_1807987_3.jpg!w690x517.jpg[/img]在下部机壳的三个定位桩及中心孔处,涂上鞋胶水(软性胶):[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721157663_714_1807987_3.jpg!w690x517.jpg[/img]将机芯对好位,装回去,压紧,搁置一天:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721158979_5830_1807987_3.jpg!w690x517.jpg[/img]待胶水凝固后,将上半部分安装回去:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721126640_6302_1807987_3.jpg!w690x517.jpg[/img]恢复正常工作:[img=,690,517]https://ng1.17img.cn/bbsfiles/images/2019/07/201907011721557919_8122_1807987_3.jpg!w690x517.jpg[/img] 化验室机械定时器这类低价值小物件,没啥高技术含量,社会上也没人愿意修理。只要掌握其装配原理,拆卸就比较容易,自己修理很简单,做到物尽其用。 根据本次的拆修过程,建议先用小电磨在定时器底部开一个孔,露出固定螺丝,卸下螺丝、取出机芯进行维修。修复后,用胶纸封口,避免暴力拆卸时损坏定时器。