应用日本京都电子公司(KEM)-QTM-500快速导热仪/导热系数测试仪_测定耐火砖的导热系数的应用资料。

智能文字提取功能测试中

上海旭风科学仪器有限公司 上海市闸北区万荣路388弄17号904室 200072Tel: 021-66581985,63172460 Fax: 021-66581986 周峰 13916388882 zhou-feng1027@163. com K EM Application Note No.TCL-07005Thermal conductivity measurement Kyoto Electronics Manufacturing Co.,Ltd. DataNo.HB00-07051 Thermal conductivity(TC) of refractory brick Summary- The purpose of this measurement is to check reproducibility of the equipment. For this,checked reproducibility with several kinds of brick, and good results was obtained withinequipment specification. -Measuring instrument- Main unit: QTM-500)Quick thermal meterProbe: PD-11 Preparation: (1) QTM was left activated for one hour untilprobe and sample were reached the setdegree. Measurement: (1) Made close contact between prove faceand sample and set the heater’s currentvolume, and measured TC. No.01一(0 1 Date :07/04/0210:36 Heater : 9.000 A2 Stable: Fine Tm : 21.3 C(T0+T2)/2 To : 18.9 C dt :4.9 C T2 :23.7 C Result : 11.630 [W/mK -Measurement parameter- -Measurement results- MEAS.PARA] Probe : PD-11 Unit :W/mK T Comp Const : off Tm Mode :(TO+T2)/2 Ts Mode: t1: 30 t2::60 TC (W/mK) Statistics 11.630 11.57111.66211.44511.633 MeanSD RSD 11.588W/mK0.0866W/mK0.747 % -Ambient condition- Ambient: 23.2 ℃ Humidity:38% Weather: Fine -Comment- Select the heater’s current volume suited to the sample. Since the ascending temperatureshould be within 5 to 25℃ in measurement.