产品介绍
The IonPac® NS1-10μm and NS1-5μm columns are packed with a neutral, macroporous, high-surface-area, ethylvinylbenzene polymer crosslinked with 55% divinylbenzene. This resin makes the NS1 resistant to solvents, acids, and bases, and permits the use of eluent from pH 0 to 14.
Ideal for separation of large molecules that carry localized charges, such as surfactants
Compatible with acids, bases, and solvent from pH 0 to 14
Can also be used for traditional polymeric reversed-phase applications
Utilize ion-pair chromatography for difficult separations
The IonPac NS1 column is the column of choice for routine ion-pair chromatography or ion-suppression separations using theAMMS® ICE 300 suppressor. The NS1 is also used with mobile phase ion chromatography (MPIC®), a technique especially suited to large molecules that carry localized charges (e.g., surfactants), as well as other ions that are not amenable to separation by ion-exchange chromatography.
The IonPac NS1 column can be used for the separation of high MW aliphatic carboxylic acids, alkyl and aromatic sulfates and sulfonates, quaternary ammonium ions, water-soluble vitamins, sulfur oxides, metal cyanide complexes, phenols, alkanolamines.
IonPac NS1 Polymeric Reversed-Phase Column Specifications | |
---|---|
Dimensions | IonPac NS1-10μm Analytical Column: 4 × 250 mm IonPac NS1-10μm Guard Column: 4 × 35 mm IonPac NS1-5μm Analytical Column: 4 × 150 mm |
Maximum operating pressure | <4000 psi (<27.6 MPa) |
Typical operating conditions | Gradient: 1.0 mL/min Recommended Operating Temperature: Ambient Recommended Flow Rate: 1.0 mL/min Typical Ion-Pairing Reagents: For anions: ammonium and tetramethyl-, tetraethyl-, tetrapropyl-, and tetrabutyl-ammonium hydroxides. For cations: hydrochloric, perchloric, and perfluorocarboxylic acids; pentane-, hexande-, heptane-, and octanesulfonic acids. |
Mobile phase compatibility | pH 0–14; 1–100% HPLC solvents |
Resin composition | Bead Diameter: 10 μm (4 × 250 mm), 5 μm (4 × 150 mm), macroporous ethylvinylbenzene/divinylbenzene resin Cross-Linking (%DVB): 55% Hydrophobicity: High |
Capacity | 60 μeq (2 × 250 mm analytical column) |
Column construction | PEEK with 10-32 threaded ferrule-style end fittings. All components are nonmetallic. |
Analytical Columns | |
---|---|
IonPac NS1-10 μm Analytical Column (4 x 250 mm) | 035321 |
IonPac NS1-5 μm Analytical Column (4 x 150 mm) | 039568 |
IonPac NS1 10 μm Analytical Column (2 x 250 mm) | SP4354 |
Guard Columns | |
---|---|
IonPac NG1-10 μm Guard Column (4 x 35 mm) | 039567 |
IonPac NG1-10 μm Guard Column (2 x 50 mm) | SP4356 |
DIONEX IONPAC NG-1 CAPILLARY GUARD (0.4X50MM) | SP6944 |