CT-04.10.Tech_.note

2019-07-31 11:06  下载量:0

资料摘要

资料下载

For personnel and equipment safety and measurement accuracy, current measurements on conductors at high voltage should be made only with a conducting shield cylinder placed inside the CT aperture. There should be a low electrical impedance connection from one end only to a reliable local ground. An inner insulating cylinder of adequate voltage isolation should be between the shield cylinder and the conductor at high voltage. Any leakage, induced or breakdown current between the high voltage conductor and the ground shield will substantially pass to local ground rather than through the signal cable to signal ground. Do not create a "current loop" by connecting the shield cylinder to ground from both ends. Current flowing in this loop will also be measured by the CT.