LaVision ParticleMaster-IMI 干涉米氏粒径测量

报价 ¥80万 - 100万

品牌

LaVision GmbH

型号

ParticleMaster-IMI

产地

欧洲德国

应用领域

共1个
no-img

Interferometric Mie Imaging

LaVision’s ParticleMaster IMI imaging system is optimized for spray investigations of smaller droplets at low and medium droplet densities. Defocused Mie scattering is used to generate a fringe pattern from each droplet with its fringe spacing related to the droplet size. This intererometric sizing method is limited to transparent and spherical droplets. The ParticleMaster IMI system is hardware compatible with LaVision’s FlowMaster 2D or Stereo PIV systems.

ParticleMaster IMI
System Features
  • autodetection with droplet location from a single camera

  • droplet size from fringe pattern analysis

  • droplet velocity derived from double frame exposures

  • velocity - size correlations, histograms, scatterplots

 
Principle of IMI droplet sizing
no-img
 

Shadowgraphy vs. Interferometric Mie Imaging

Both, ParticleMaster Shadow and IMI are based on imaging techniques to measure particle size. Their limit to smaller size is determined by the nature of light. While the shadowgraphy technique uses a direct image, IMI is based on coherent light scattering and allows to extend the range of detectable droplets to even smaller sizes.

ParticleMaster
Application Matrix

no-img