Kyoto Electronics Manufacturing Co.,Ltd.DataNo.HB00-07051 KEM Application Note No.TCL-07005Thermal conductivity measurement Thermal conductivity(TC) of refractory brick -Summary- The purpose of this measurement is to check reproducibility of the equipment. For this,checked reproducibility with several kinds of brick, and good results was obtained withinequipment specification. -Measuring instrument- Main unit: QTM-500 Quick thermal meterProbe: PD-11 (1) QTM was left activated for one hour untilprobe and sample were reached the set degree. Measurement: (1) Made close contact between prove faceand sample and set the heater’s currentvolume, and measured TC. No.01-01 Date :07/04/02 10:36 Heater:9.000 A2 Stable:Fine Tm :21.3 C(T0+T2)/2 To :18.9 C dt :4.9 C T2 :23.7C Result : 11.630 [\W/mK -Measurement parameter- -Measurement results- [MEAS.PARA1 Unit Probe :PD-11:W/mK TComp Const : off Tm Mode:(TO+T2)/2 Ts Mode : t1:30 t2: 60 TC (W/mK) Statistics 11.63011.57111.66211.445 11.633 MeanSD RSD 11.588W/mK0.0866W/mK0.747% -Ambient condition- Ambient:23.2 C Humidity: 38% Weather: Fine -Comment- Select the heater’s current volume suited to the sample. Since the ascending temperatureshould be within 5 to 25℃ in measurement. 京都电子工业株式会社-可睦电子(上海)商贸有限公司上海市漕溪路222号航天大厦北楼902室电话:021-54488867 电邮: kemu-kem@163. com网址: http://kem. instrument.com.cn