Test Method For Particle Size Reference Material LUDOX TM On LA-950

2016/12/13   下载量: 3

方案摘要

方案下载
应用领域 纺织/印染
检测样本
检测项目
参考标准 0

LUDOX silica is a well known and characterized colloid that has been studied using various particle size analysis techniques including acoustic spectroscopy, laser diffraction, and dynamic light scattering. One such product, LUDOX TM-50, is a 50 wt% suspension of colloidal silica in water. It can be used as a “real world” reference material to test the performance of laser diffraction particle size analyzers. As with many colloids, good dispersion is critical for LUDOX TM to meet specifications. The importance of using a simple salt solution to dilute this material is shown below. LUDOX TM-50 has been tested and incorporated as a particle size reference material for the LA-950. ?

方案下载
配置单
方案详情

LUDOX silica is a well known and characterized colloid that has been studied using various particle size analysis techniques including acoustic spectroscopy, laser diffraction, and dynamic light scattering. One such product, LUDOX TM-50, is a 50 wt% suspension of colloidal silica in water. It can be used as a “real world” reference material to test the performance of laser diffraction particle size analyzers. As with many colloids, good dispersion is critical for LUDOX TM to meet specifications. The importance of using a simple salt solution to dilute this material is shown below. LUDOX TM-50 has been tested and incorporated as a particle size reference material for the LA-950.


上一篇 荧光光谱+有机掺杂材料+构建
下一篇 椭圆偏振光谱仪的标准应用

文献贡献者

HORIBA(中国)
白金会员第22年
查看全部资料
相关仪器 更多
相关方案
更多

相关产品

当前位置: HORIBA 方案 Test Method For Particle Size Reference Material LUDOX TM On LA-950

关注

拨打电话

留言咨询